<?xml version="1.0" standalone="yes"?>
<?xml-stylesheet type="text/xsl" href="..\..\XSLTemplates\ErrorsStyleSheet.xsl"?>
<Error>
<ErrorNumber>22117</ErrorNumber>
<ErrorFamily>Autoloader 2</ErrorFamily>
<ErrorMessage>
ALE2: Failed to Find focus, please have media reflectivity setup checked
</ErrorMessage>
<AppliesTo>
<UnnumberedList>
<ListItem><Text>ThermoFlex</Text></ListItem>
</UnnumberedList>
</AppliesTo>
<ServiceInformation>
<DescriptionText><text>
Revisit the surface reflectivity of the media in use.
Use the diagnostic stage moveto command and the focus cal reflect command to
reset the sr value to a correct value.
</text></DescriptionText>
</ServiceInformation>
<OperatorAction>
<DescriptionText><text>
Retry &amp; Call Service
</text></DescriptionText>
</OperatorAction>
<Cause>
The head stage has tried far too many moves in its attempt to find focus and
has timed out.  This is almost always due to an incorrect surface reflectivity
causing the values from the head to confuse the zeroing algorithm.
</Cause>
<SeeAlso>
<UnnumberedList>
<Link><LinkAddress>../commands/focus_cal.xml</LinkAddress><LinkTitle>Command: focus cal</LinkTitle></Link>
<Link><LinkAddress>../nvs/media_cal_sr.xml</LinkAddress><LinkTitle>NVS: media sr</LinkTitle></Link>
</UnnumberedList>
</SeeAlso>
<OtherLinks>
<Link><LinkAddress>..\..\ShowAllErrors.htm</LinkAddress><LinkTitle>Firmware 0 Error Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllCommands.htm</LinkAddress><LinkTitle>Firmware 0 Command Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllNVS.htm</LinkAddress><LinkTitle>Firmware 0 NVS Index</LinkTitle></Link>
</OtherLinks>
<Updated>
Updated: Wed Jul 16 19:33:56 2008 PST
</Updated>
<Links>
</Links>
</Error>